Book a service
About this product

Description

Gram BioPlus ER660W provides temperature-controlled storage for medicines and laboratory materials within a −2 to +20 °C setting range. Its 660-litre gross volume gives 484 litres of usable net space. The wide cabinet is 815 mm across and 756 mm deep.

A ventilated BioLine air system distributes cooling through the stainless steel interior. The Gram controller includes an E-sensor, a voltage-free contact and temperature adjustment. Audible and visual alarms signal a temperature deviation or an open door.

Defrosting is automatic, defrost water is evaporated and the door is self-closing. Solid and glass doors are available, with white or stainless steel exterior finishes. Shelves and drawers are selected to suit the stored material; 27 shelf or 13 drawer module positions do not mean that these quantities are included.

Energy use for the solid-door R290 model is 1.28 kWh per day; glass-door and dual-compressor figures are listed separately in the specifications.

Technical data

Specifications

Type
Medical refrigerator
Manufacturer
GRAM
Model name
BioPlus ER660W
Gross volume; litres
660
Net volume; litres
484
Temperature adjustment range
-2 °C to 20 °C
Ambient conditions
Solid door: +10 to +43 °C; glass door: +10 to +38 °C
Interior finish
Stainless steel
Exterior finish
White lacquered steel or stainless steel
Door / drawer
Self-closing solid door; optional glass door
External dimensions (W×D×H); mm
815 x 756 x 2025/2275 mm
Gross and net weight; kg
Unpackaged net weight without optional fittings: 152 kg
Shelf and drawer module positions
27 shelf or 13 drawer module positions; fittings selected separately
Insulation
60 mm polyurethane with HFC-free cyclopentane propellant
Thermal transmittance
Solid door 0.31 W/m²K; glass door 0.49 W/m²K
Temperature controller
Gram controller: voltage-free contact, E-sensor and temperature adjustment; dry-cooling function
Alarms
Acoustic and visual, High/Low temperature alarms and door alarm
Sensor access port
1 port, Ø 24.5 mm
Cooling
BioLine ventilated air distribution system
Defrosting
Automatic smart defrost with re-evaporation of defrost water
Refrigerant
R290
Refrigerant charge
102 g
Cooling capacity
R290 389 W (−10 °C)
Energy consumption; kWh/24h
Solid door 1.28; glass door 1.78; dual-compressor solid door 2.20. Dual-compressor glass-door figure not provided.
Heat rejection
  • Solid door 315.5 W
  • Glass door 320.46 W
  • Dual-compressor solid door 549.25 W. Dual-compressor glass-door figure not provided.
  • 23 °C ±1 K, 100% operation
Electrical connection
230 V, 50 Hz
Electrical power; W
Single compressor 335; dual compressor 587
Starting current
9.7 A
Noise level; dBA
48.5
Feet / castors
Legs
Optional accessories
Optional castors or plinth mounting
Warranty
24 months
Similar products

See also

Buying information

Delivery, warranty and ordering

Medical refrigerator BioPlus ER660W

Request a quote or advice

We can confirm price and availability and help you choose a suitable model.